Refer Standards
ISO 7886-1:2017 Sterile hypodermic syringes for single use - Part 1: Syringes for manual use.
ISO 80369-7:2016 Small -bore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications.
USP <85> Bacterial Endotoxins Test.
USP <788> Particulate Matter in Injections.
USP <791> PH.
USP <71> Sterility Tests.
USP <231> Heavy Metals
USP <661> Plastic Packaging Systems and Their Materials of Construction.
ISO 10993-18: 2020 Biological evaluation of medical devices - Part 18: Chemical characterization of medical device materials within a risk management process.
Suitable for R & D / F & D purpose & small batch size production
Suitable for Nested Pre-fill Syringes & Nested Stopperes
Manual transfer of Tub from Filling to Stoppering Unit
21CFR Part 11 Compliance
Suitable for aqueous & viscous & visco - elastics drugs
Filling Accuracy @ (+/-) 0.02ml in 0.2ml Fill
Fill Volume Spectrum : 0.2ml to 10ml (Change Part Supported)
Sterile Hypodermic Syringes with Needle for Single Use
US$ 0.5 / Carton Get Freight Cost Calculate Margin
10000 Cartons Minimum Order : 20 - 90 days
Product Details
Model Number : ZS-05- H-SY 5ML-#3409
Brand Name : Winhealth
Origin : China (mainland)
Small Orders : Accepted
Key Specifications/Special Features:
Model No. ZS-01-H-SY 1ML, 3 ML, 5ML
Brand Name: Winhealth
Origin: China
CE Approval
Fast Delivery for Any Quantity
Sterilization Independent Packaging
Using Independent packaging ,after ethylene oxide sterilization, open ready tp use.
All Sterile EO Processing with Single Blister Pack with Needle Pre-attached
Different Sizes: 1ML, 3ML, 5ML
Shipping Information:
FOB Port: Shanghai Lead Time : 20 - 90 days
HTS Code : 9018.31.00
Dimensions per Unit : 15 X 3 X 3 Centimeters
Weight per Unit : 7 Grams Units per Export
Carton : 1600 Export Carton
Dimensions L/W/H: 48.5 X 40.5 X 45 Centimeters Export Carton
Weight : 8.5 Kilograms
Main Export Markets:
Asia
asia
Central/South America
Eastern Europe
Mid East/Africa
North America
Western Europe
Any third-party trademarks or images shown here are for reference purposes only. We are not authorized to sell any items bearing such trademarks.
Payment Details:
Payment Method : Telegraphic Transfer in Advance (Advance TT, T/T)
Injection in vail (1,3,5 MIU/ml);
Treat chronic hepatitis B, chronic hepatitis C, tumor and other viral infective disease with a broad spectrum of anti-tumor and immuno-regulating functions.
General Characteristics
Single-use
Sterile
Including a mechanism to prevent reuse
The syringe and needle shall be passively and automatically rendered unusable by the delivery of the intended fixed dose. And/ Or
The auto-disable feature is automatically activated and remains effective from the time that the injection is commenced. And / Or
The auto-disable feature is automatically activated on completion of the delivery of the intended fixed dose.
Type and application
Non-pre-filled, disposable.
To administer DTP, DT, tetanus toxoid and diluted freeze dried measles etc for intramuscular and sub-cutaneous injection.
Syringes with permanently attached needle
Material
Polypropylene with or without stainless steel
Syringe size with pre-set volume and single marking
0.5ml for immunization (+20% to allow for removal of air)
Graduation and accuracy
Two markings only: 0 and 0.5ml.
The difference between the maximum graduated volume and the expelled volume of distilled water shall not be greater than �± 25 micro-liters.
Needle for Immunization
· Diameter : 23G
Length : 25mm (1)
Stainless steel needle tubing for the manufacture of medical devices
Needle shall confirm to IS 10654 of 2002 (third revision) equivalent ISO 7864 of 1993
Needle fixing
Fixed capped, needle (23 G X 25 mm). Fixed needles should withstand an axial pulling force of 34N without separating from the syringe. If accidental needle stick protection devices are included in the design of the syringes or needle, they should not compromise the ease of use or the performance of the syringe.
Leakage :
Leakage of distilled water at a pressure of 0.3 mPa (3 atmospheres) for six minutes shall not be greater that 50 micro-liters.
Dead space
The maximum dead-space shall be 50 micro-liters.
Self destruction
The syringe shall be automatically rendered unusable when it has delivered a 0.5 ml dose of vaccine or distilled water.
Filling
The syringe is filled by inserting the needle through the septum of a sealed vaccine vial.